De-identification policy¶
This page defines exactly what each de-identification level does, how that maps to
DICOM PS3.15 Annex E (Basic Application Level Confidentiality Profile), and what is
not covered. It is the reference for src/dicomforge/deidentify.py; the tests in
tests/test_deid_policy.py check each rule.
Scope
De-identification here is header-only risk reduction. It is not a HIPAA, GDPR or other regulatory certification, and it has not been clinically validated. Pixel data is never analysed or modified. Series whose metadata signals burned-in annotation are rejected; text burned into any other image survives. Review output before it leaves a controlled environment.
Where it applies¶
The pipeline de-identifies the DICOM datasets in memory, then converts them to NRRD/NIfTI. Source files are never modified, and no de-identified DICOM is written. The converted volume contains no DICOM header fields at all: only voxels and geometry. The header policy below therefore governs:
- the in-memory datasets, for API users
- everything derived from them: the audit record (
PipelineResult), QC report and logs
The voxels in the output are always identical to the source pixels, at every level.
Rules applied at every level¶
| Category | Action | PS3.15 action |
|---|---|---|
| Person names: every PN element at any depth (patient, physicians, operators…), also when stored with VR UN | removed | X |
| Patient identifiers and demographics: birth date/time (also in alternative calendars), address, phone and telecom, other IDs and names, issuer of patient ID, ethnic group, occupation, religion, residence and current location, insurance, medical alerts, allergies | removed | X |
Contact details of any person (PersonAddress, PersonTelephoneNumbers, PersonTelecomInformation), at any depth |
removed | X |
Overlays (60xx) and curves (50xx): whole repeating groups, including OverlayData bitmaps and overlay descriptions/labels |
removed | X |
| Public tags outside the DICOM dictionary (cannot be interpreted, so cannot be checked) | removed | (conservative) |
| Institution name, address, department, institution code sequence | removed | X |
Identifying sequences: OtherPatientIDsSequence, ReferencedPatientSequence, RequestAttributesSequence, ReferencedStudySequence, ReferencedPerformedProcedureStepSequence, physician/operator identification sequences |
removed with all contents | X |
| Order and visit IDs and descriptions: requested/scheduled/performed procedure step IDs and descriptions, admission, service episode | removed | X |
Free-text comments: ImageComments, StudyComments, AcquisitionComments, PatientComments, AdditionalPatientHistory, InterpretationText, request comments |
removed | X |
AccessionNumber, StudyID |
kept, zero-length | Z |
| All private tags (creators and data), at any depth | removed | X (no "Retain Safe Private" option) |
| Instance UIDs at any depth and in file meta (Study, Series, SOP Instance, Frame of Reference, referenced instances…) | replaced with 2.25.<n> |
U |
PatientID, wherever it appears |
replaced with a keyed pseudonym DF-… |
(pseudonymisation) |
| Remaining text (LO/SH/ST/LT/UT/UC) containing one of this dataset's own identifiers | blanked | (Clean Descriptors, limited) |
Rules apply recursively. Every sequence item is processed with the same rules, however deeply nested.
UID replacement uses HMAC-SHA256 under the pseudonymisation secret, so it is:
- consistent: the same original UID always maps to the same replacement. Slices of a series keep one Series UID, and references between instances still resolve.
- unlinkable without the secret.
UIDs defined by the DICOM standard (1.2.840.10008.*) are kept. So are class,
transfer-syntax and implementation UIDs (e.g. SOPClassUID, ReferencedSOPClassUID,
TransferSyntaxUID), because they identify a kind of object, not a patient.
Known-identifier screening collects, before scrubbing:
- every person-name component of 3+ characters
- the values of PatientID, other patient IDs, issuer, AccessionNumber, StudyID, birth date, requested-procedure and admission IDs, medical record locator and institution name
- every original instance UID (so a UID quoted in text is caught)
- at moderate and strict, every date value (
YYYYMMDD), so a date quoted in text doesn't survive when dates are being removed
It then blanks any remaining text element that contains one of them, case-insensitively.
For example, StudyDescription = "BRAIN FOLLOWUP TEST-12345" is blanked, while
SeriesDescription = "T1 AXIAL" is kept.
Provenance:
PatientIdentityRemoved = NO. DICOM defines this attribute as covering the attributes and the pixel data. dicom-forge never verifies pixel content, so it can't assertYES, even when the headers are fully de-identified.DeidentificationMethod = "dicom-forge <level>: header de-identified; pixels unverified"records what was actually done.BurnedInAnnotationis left exactly as the source declared it.LongitudinalTemporalInformationModifiedis set toUNMODIFIED(basic) orREMOVED(moderate, strict).- No
DeidentificationMethodCodeSequenceis written. The Basic Profile code (113100) would claim conformance that this subset doesn't meet.
Per-level differences¶
| basic | moderate | strict | |
|---|---|---|---|
| Dates and times (DA/DT/TM), at any depth | kept (Retain Longitudinal Temporal Information) | blanked | blanked |
Device identifiers: StationName, DeviceSerialNumber, DetectorID, GantryID, and every AE title (including file-meta source/sending AE titles) |
kept (unless screened) | removed | removed |
Patient characteristics: PatientAge, PatientSex, PatientSize, PatientWeight |
kept (Retain Patient Characteristics) | kept | age/size/weight removed; sex blanked |
| All remaining free text (LO/SH/ST/LT/UT/UC/AS) outside the geometry safe-list | screened only | screened only | blanked |
PatientID when pseudonymisation is disabled |
raw ID kept (unsafe opt-out) | raw ID kept (unsafe opt-out) | blanked |
The STRICT safe-list (never blanked): Modality, Rows, Columns, PixelSpacing,
SliceThickness, ImagePositionPatient, ImageOrientationPatient, RescaleSlope,
RescaleIntercept, PixelData. Numeric (DS/IS/US…) and coded (CS) attributes are not
blanked by the strict text rule. That keeps geometry and conversion identical across
levels.
What each level guarantees, and what it does not¶
- basic: no names, direct patient identifiers, institution, accession number, private tags or original UIDs remain anywhere in the header. Dates, device names, descriptions without known identifiers, and age/sex/size/weight remain. Suitable only where dates and device information are acceptable to disclose.
- moderate (default): basic, plus no dates/times and no device identifiers. Age, sex, size and weight remain, as do descriptions without known identifiers (e.g. protocol names).
- strict: moderate, plus no patient characteristics and no free text beyond coded/numeric attributes and the safe-list.
At no level are pixel data, overlays (60xx) or encapsulated documents processed.
Burned-in pixel annotation¶
dicom-forge does not detect, remove or mask text in pixel data, and makes no claim
of automatic pixel de-identification. Instead, before anything is de-identified or
written, it assesses every instance from metadata only
(dicomforge.pixel_privacy.assess_burned_in_risk) and flags the series if any
instance shows a risk signal:
| Reason code | Signal |
|---|---|
burned_in_annotation_yes |
BurnedInAnnotation (0028,0301) is YES |
burned_in_annotation_unrecognised |
BurnedInAnnotation holds anything other than YES/NO |
secondary_capture |
a Secondary Capture SOP class (1.2.840.10008.5.1.4.1.1.7*), or ConversionType present |
high_risk_modality |
Modality US, SC, OT, XC, ES, GM or SM, where on-image text and labels are routine |
colour_pixels |
PhotometricInterpretation not MONOCHROME1/2 (screenshots, rendered captures) |
overlay_in_pixel_data |
an overlay (60xx) stored in the pixel bits instead of its own OverlayData plane |
Policy (PipelineConfig.burned_in_policy, CLI --allow-burned-in):
reject(default): a flagged series raisesBurnedInAnnotationErrorbefore de-identification and before any file is written. The CLI exits with code3. One flagged instance rejects the whole series. This applies at every level, and also with de-identification disabled.allow: the flagged series is processed, a warning is logged, and the audit record keepsstatus = "flagged". Use it only after a person has reviewed the images.
The rejection message and logs carry only reason codes and counts: no paths, UIDs, names or attribute values.
Unflagged is not clean. A series without any signal, including one that declares
BurnedInAnnotation = NO, is processed with status = "unverified". Every
PipelineResult.pixel_privacy record has pixel_content_verified = false and
pixels_modified = false. The record also counts instances per indicator value
(YES / NO / absent / other).
What the burned-in policy cannot catch¶
- Text in CT, MR, PET or other "low-risk" images with no metadata signal. For
example, a scanner or post-processing step that stamped the patient name into an MR
image without setting
BurnedInAnnotation. This is a tested, documented limitation (test_limitation_unflagged_burned_in_text_is_not_detected). - Incorrect source declarations:
BurnedInAnnotation = NOon an image that has text, or a modality code that understates the risk. - Derived and reformatted images (
ImageTypeDERIVED/SECONDARY, MPRs, MIPs) are not flagged on that basis, even though some vendors annotate them. - Non-text identifiers in pixels: faces in 3-D surface renderings of head CT/MR, facial features recoverable from volumes (defacing is not performed), and identifiable jewellery, implants or tattoos.
- Multi-frame, colour and compressed objects are flagged or handled only by the signals above. No pixel decoding is used to look for text.
Audit record¶
PipelineResult.deidentification covers the whole series:
removed_tags,blanked_tagsandreplaced_uidsare unions over every instance, with nested paths writtenSequence[].Keywordinstancesis the number of instances coveredremoved_private_tagsis the total countpseudonymised_patient_idslists every pseudonym seenpseudonymised_patient_idis set only when the series maps to exactly one
The metadata in the record is read after de-identification, its UIDs are the
replacement UIDs, and source_path is always None. See
Concepts → Audit records and logs.
Converted volumes say what was (not) done¶
Every NRRD written by the pipeline carries these header fields (key:=value). NIfTI has
no place for them; the audit record is the source of truth there.
| Field | Values |
|---|---|
dicomforge_version |
engine version, e.g. 0.2.0 |
dicomforge_header_deidentification |
basic / moderate / strict (applied to the source DICOM headers), or none with --no-deid |
dicomforge_pixel_data |
unverified for every normal run; flagged-burned-in-risk-allowed when the series was flagged and processed only because of the explicit override. Never "clean". |
The Slicer extension also sets these values as volume-node attributes
(DicomForge.PixelData, DicomForge.HeaderDeidentification, DicomForge.EngineVersion).
It adds _UNSAFE_BURNED_IN to the name of a node loaded through the override.
What this protects against, and what it does not¶
Protects against (header data, tested with fictional identifiers at every level): names, patient IDs (pseudonymised), birth dates, addresses and contact details, institution and staff identifiers, accession/order/visit identifiers, private vendor tags, overlays and curves, unknown public tags, original UIDs (consistently replaced), identifiers nested in sequences, and the record's own identifiers quoted in free text. It also refuses series whose metadata signals burned-in pixel text, unless explicitly overridden. Identifiers are kept out of logs, audit records, CLI output and error messages.
Does not protect against:
- identifying text in pixel data without a metadata signal
- facial features in head CT/MR
- identifiers in free text that are not this record's own
- re-identification from the quasi-identifiers it deliberately keeps (age, sex, size,
weight; dates at
basic) - anyone holding the pseudonymisation secret
- attributes outside this documented PS3.15 subset
It is not certified for HIPAA, GDPR or any other regulation, and it is not a validated anonymization tool.
Known limitations¶
- Burned-in annotations in pixel data are never removed. They're caught only when metadata signals them (see above). Overlay and curve groups are removed from the header, but encapsulated documents (e.g. PDF) and structured-report content beyond person names are not de-identified.
- Free-text screening is not general PHI detection. It finds only this dataset's
own identifiers, and only tokens of 3+ characters, matched literally. A description
naming a different person, a 2-letter surname, or an ID written differently
(
TEST12345forTEST-12345) will not be caught. Usestrictwhen free text cannot be trusted. - Quasi-identifier risk is not measured. No k-anonymity or similar analysis is done. Small cohorts, rare conditions or unusual devices can make records re-identifiable even with every listed identifier removed.
- Dates are blanked, not shifted. Intervals between studies are lost at moderate and strict.
- Retained characteristics (age, sex, size, weight at basic/moderate; dates at basic) are quasi-identifiers. Combined, they can single out individuals in small populations.
- Code sequences that are not in the removal list are kept, with only their text screened. Coded values (CS) are never blanked.
- The pseudonymisation secret is the single point of trust. Anyone holding it can re-link pseudonyms and replacement UIDs by testing guesses.